LiveActive security incident?Get immediate response
CVE Record

CVE-2021-33883: A Cleartext Transmission of Sensitive Information vulnerability in B.

A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.

MediumCVSS 5.9Not KEV-listedUpdated
Glexia's TakeAutomated analysismoderate

Security readout for executives and security teams

Plain-English summary

CVE-2021-33883 means some B. Braun SpaceCom2 versions sent sensitive pump configuration information without encryption. Someone with access to observe the relevant network traffic could learn internal pump configuration values. The CVE is not listed in KEV, so the provided sources do not show confirmed active exploitation.

Executive priority

Treat this as a targeted medical-device remediation item. It is not presented as actively exploited, but hospitals using affected SpaceCom2 systems should prioritize inventory, segmentation, and vendor-confirmed updates because exposed pump configuration data can affect clinical technology risk.

Technical view

B. Braun SpaceCom2 before 012U000062 is described as transmitting sensitive information in cleartext. The CVSS 3.1 score is 5.9 with network attack vector, no privileges, no user interaction, high attack complexity, and high integrity impact. The source metadata does not provide complete affected CPEs.

Likely exposure

Exposure is most likely in healthcare environments using B. Braun SpaceCom2 versions earlier than 012U000062 on networks where unauthorized users or compromised hosts can observe device traffic.

Exploitation context

The described attack depends on network visibility to the relevant traffic. The sources do not provide evidence of exploitation in the wild, and KEV status is false. The main concern is sensitive medical-device configuration exposure that could support broader operational risk.

Researcher notes

Evidence is limited to CVE metadata and referenced vendor/research pages. The CVE record names SpaceCom2 before 012U000062, but the affected product fields are listed as n/a. Do not assume other B. Braun products are affected from this bundle alone.

Mitigation direction

  • Inventory B. Braun SpaceCom2 deployments and firmware versions.
  • Check B. Braun guidance for updates to 012U000062 or later.
  • Segment medical-device networks from user and guest networks.
  • Limit who can access or observe pump-related network traffic.
  • Coordinate remediation windows with clinical engineering and patient-safety stakeholders.

Validation and detection

  • Confirm no SpaceCom2 device is running below 012U000062.
  • Review network segmentation around infusion pump infrastructure.
  • Check whether pump-related traffic is exposed to non-clinical network segments.
  • Verify vendor remediation guidance has been applied and documented.
Prepared
Confidence
medium
Sources
4

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cve · low confidence lookup

CVE-2021-33883 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Medium
CVSS
5.9 (3.1)
Known Exploited
No
Published

Vector: CVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:U/UI:N

Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

1CVSS vectors
0Timeline events
0ADP providers
3Source links

CVSS vector scores

1 official score

We collect every scored CVSS vector available in the official CNA and ADP containers. When more than one version is present, the table keeps the source vectors side by side instead of collapsing them into the highest score.

ScoreVersionSeverityVectorExploitImpactSource
5.9CVSS 3.1MediumCVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:U/UI:N2.23.6Primary CVE score

Vulnerability scoring details

Base CVSS 3.1 score

5.9Medium
CVSS 3.1 vector shape for CVE-2021-33883Attack VectorAttack ComplexityPrivileges RequiredUser InteractionScopeConfidentiality ImpactIntegrity ImpactAvailability Impact

Vector: CVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:U/UI:N

Attack Vector
NetworkAdjacentLocalPhysical
Attack Complexity
LowHigh
Privileges Required
NoneLowHigh
User Interaction
NoneRequired
Scope
ChangedUnchanged
Confidentiality Impact
HighLowNone
Integrity Impact
HighLowNone
Availability Impact
HighLowNone
Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
n/an/an/aListed
Weakness

CWE details

No CWE listed

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.