LiveActive security incident?Get immediate response
CVE Record

CVE-2021-33882: A Missing Authentication for Critical Function vulnerability in B.

A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.

MediumCVSS 6.8Not KEV-listedUpdated
Glexia's TakeAutomated analysismoderate

Security readout for executives and security teams

Plain-English summary

This issue affects B. Braun SpaceCom2 versions before 012U000062. A remote attacker could change device configuration because some proprietary networking commands lacked authentication. The public bundle does not show active exploitation, but unauthorized configuration changes in infusion infrastructure can create clinical and operational risk.

Executive priority

Treat as a healthcare operational risk requiring inventory and remediation planning, not an internet-scale emergency. Prioritize affected clinical sites, verify fixed versions, and coordinate changes with biomedical engineering and the vendor.

Technical view

CVE-2021-33882 is missing authentication for a critical function in SpaceCom2 before 012U000062. The CVSS 3.1 vector is network-accessible, no privileges, no user interaction, high attack complexity, changed scope, and high integrity impact only. The bundle lists no CWE and incomplete affected-product metadata.

Likely exposure

Exposure is most likely in healthcare networks running B. Braun SpaceCom2 earlier than 012U000062. Risk increases if devices are reachable from broad internal networks or poorly segmented environments. The source bundle’s structured affected list is incomplete.

Exploitation context

No KEV listing or provided source states active exploitation. The CVE describes remote reconfiguration from an unknown source using unauthenticated proprietary networking commands, but high attack complexity limits broad opportunistic risk based on the supplied CVSS vector.

Researcher notes

The description is specific, but the structured affected data is n/a and no CWE is supplied. Avoid assuming other B. Braun products are affected. Use the CVE record, CVE JSON, vendor materials, and McAfee analysis to confirm exact scope.

Mitigation direction

  • Identify all B. Braun SpaceCom2 deployments and versions.
  • Prioritize upgrades to 012U000062 or later where vendor guidance confirms applicability.
  • Check B. Braun guidance before applying device or clinical-network changes.
  • Restrict device management access to trusted clinical engineering networks.
  • Review segmentation and monitoring around infusion pump infrastructure.

Validation and detection

  • Confirm whether any SpaceCom2 systems run versions before 012U000062.
  • Validate device management paths are not broadly reachable across the network.
  • Compare device configuration against approved clinical engineering baselines.
  • Review vendor documentation for the exact fixed release and deployment process.
  • Document exceptions where upgrades require clinical or vendor coordination.
Prepared
Confidence
medium
Sources
4

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cve · low confidence lookup

CVE-2021-33882 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Medium
CVSS
6.8 (3.1)
Known Exploited
No
Published

Vector: CVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:C/UI:N

Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

1CVSS vectors
0Timeline events
0ADP providers
3Source links

CVSS vector scores

1 official score

We collect every scored CVSS vector available in the official CNA and ADP containers. When more than one version is present, the table keeps the source vectors side by side instead of collapsing them into the highest score.

ScoreVersionSeverityVectorExploitImpactSource
6.8CVSS 3.1MediumCVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:C/UI:N2.24Primary CVE score

Vulnerability scoring details

Base CVSS 3.1 score

6.8Medium
CVSS 3.1 vector shape for CVE-2021-33882Attack VectorAttack ComplexityPrivileges RequiredUser InteractionScopeConfidentiality ImpactIntegrity ImpactAvailability Impact

Vector: CVSS:3.1/AC:H/AV:N/A:N/C:N/I:H/PR:N/S:C/UI:N

Attack Vector
NetworkAdjacentLocalPhysical
Attack Complexity
LowHigh
Privileges Required
NoneLowHigh
User Interaction
NoneRequired
Scope
ChangedUnchanged
Confidentiality Impact
HighLowNone
Integrity Impact
HighLowNone
Availability Impact
HighLowNone
Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
n/an/an/aListed
Weakness

CWE details

No CWE listed

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.