LiveActive security incident?Get immediate response
CVE Record

CVE-2020-25156: B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.

HighCVSS 7.2Not KEV-listedUpdated
Glexia's TakeAutomated analysishigh

Security readout for executives and security teams

Plain-English summary

This vulnerability affects B. Braun infusion-related communication modules. Leftover debug functionality can let an attacker with required cryptographic material gain root-level control of the device, creating serious confidentiality, integrity, and availability risk in clinical environments.

Executive priority

Treat this as a high-priority medical device exposure review, especially where affected communication modules are network reachable. The main business risk is unauthorized root access to clinical device infrastructure, not confirmed mass exploitation.

Technical view

CVE-2020-25156 is CWE-489 active debug code in SpaceCom L8/U61 and Data module compactplus A10/A11 and earlier, with Battery Pack SP with Wi-Fi also listed. CVSS 3.1 is 7.2, network reachable, low complexity, high privileges required, no user interaction, and high CIA impact.

Likely exposure

Exposure is most relevant to healthcare organizations using the listed B. Braun SpaceCom, Battery Pack SP with Wi-Fi, or Data module compactplus versions on reachable clinical or operational networks.

Exploitation context

The provided sources do not show CISA KEV listing or active exploitation. Exploitation requires possession of cryptographic material, but successful access would be root-level and could compromise device control and data.

Researcher notes

Evidence is limited to the CVE description and advisory references in the provided bundle. Do not assume affected versions beyond those listed. High privileges and cryptographic material are required, but impact is high after successful access.

Mitigation direction

  • Check B. Braun and CISA advisories for current vendor remediation guidance.
  • Inventory SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus deployments.
  • Prioritize remediation for devices reachable from shared or less-trusted networks.
  • Restrict network access to affected devices pending vendor-approved fixes or controls.
  • Coordinate changes with clinical engineering and patient-safety owners.

Validation and detection

  • Identify deployed product models and firmware or module versions.
  • Confirm whether SpaceCom L8/U61 or Data module compactplus A10/A11 or earlier are present.
  • Review network paths to determine whether affected devices are reachable.
  • Check security monitoring for unauthorized administrative or root-level access indicators.
  • Document compensating controls and vendor remediation status for each affected device group.
Prepared
Confidence
medium
Sources
4

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-489: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2020-25156 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
High
CVSS
7.2 (3.1)
Known Exploited
No
Published

Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

1CVSS vectors
0Timeline events
0ADP providers
3Source links

CVSS vector scores

1 official score

We collect every scored CVSS vector available in the official CNA and ADP containers. When more than one version is present, the table keeps the source vectors side by side instead of collapsing them into the highest score.

ScoreVersionSeverityVectorExploitImpactSource
7.2CVSS 3.1HighCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H1.25.9Primary CVE score

Vulnerability scoring details

Base CVSS 3.1 score

7.2High
CVSS 3.1 vector shape for CVE-2020-25156Attack VectorAttack ComplexityPrivileges RequiredUser InteractionScopeConfidentiality ImpactIntegrity ImpactAvailability Impact

Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

Attack Vector
NetworkAdjacentLocalPhysical
Attack Complexity
LowHigh
Privileges Required
NoneLowHigh
User Interaction
NoneRequired
Scope
ChangedUnchanged
Confidentiality Impact
HighLowNone
Integrity Impact
HighLowNone
Availability Impact
HighLowNone
Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
B. Braun Melsungen AGSpaceComunspecified, unspecifiedListed
B. Braun Melsungen AGBattery pack with Wi-Fiunspecified, unspecifiedListed
B. Braun Melsungen AGData module compactplusA10, A11Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-489 · source CWE mapping

Active Debug Code

Active Debug Code represents a recurring weakness pattern that can create exploitable paths when design, validation, or implementation controls are missing.