LiveActive security incident?Get immediate response
CVE Record

CVE-2020-25154: B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites.

MediumCVSS 5.4Not KEV-listedUpdated
Glexia's TakeAutomated analysismoderate

Security readout for executives and security teams

Plain-English summary

This vulnerability lets an attacker make the device administrative interface send a user to a malicious website. The likely business risk is phishing or credential theft around medical device administration, not direct device takeover from the evidence provided.

Executive priority

Handle as a moderate medical technology security issue. Prioritize inventory and access restriction, then follow vendor guidance for remediation. Escalate if affected admin interfaces are broadly reachable or used by many staff.

Technical view

CVE-2020-25154 is a CWE-601 open redirect in B. Braun SpaceCom L81/U61 and earlier and Data module compactplus A10/A11. CVSS 3.1 is 5.4: network reachable, low complexity, no privileges, user interaction required, with low confidentiality and integrity impact.

Likely exposure

Exposure is most relevant where affected B. Braun device administrative interfaces are reachable by users or administrators. Battery Pack SP with Wi-Fi is named, but version detail is incomplete in the provided bundle.

Exploitation context

The source bundle does not show CISA KEV listing or confirmed active exploitation. Practical abuse would require tricking a user into following a crafted path involving the administrative interface.

Researcher notes

Evidence supports open redirect impact only. Do not infer code execution, therapy manipulation, or availability impact from this bundle. CVSS indicates user interaction is required and no privileges are required.

Mitigation direction

  • Check B. Braun and CISA advisories for confirmed remediation or firmware guidance.
  • Inventory SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus deployments.
  • Restrict administrative interfaces to trusted management networks and authorized administrators.
  • Treat suspicious redirect reports involving device administration as phishing indicators.

Validation and detection

  • Confirm device models and firmware versions against the affected product list.
  • Verify whether administrative interfaces are reachable from non-management networks.
  • Review available web or access logs for unusual redirect-related activity.
  • Document any vendor-confirmed fixed versions or compensating controls applied.
Prepared
Confidence
medium
Sources
4

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · medium confidence lookup

CWE-601: User-session and phishing behavior lookup

Client-side and session-facing weaknesses should be reviewed alongside initial-access and user-execution behaviors. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2020-25154 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Medium
CVSS
5.4 (3.1)
Known Exploited
No
Published

Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N

Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

1CVSS vectors
0Timeline events
0ADP providers
3Source links

CVSS vector scores

1 official score

We collect every scored CVSS vector available in the official CNA and ADP containers. When more than one version is present, the table keeps the source vectors side by side instead of collapsing them into the highest score.

ScoreVersionSeverityVectorExploitImpactSource
5.4CVSS 3.1MediumCVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N2.82.5Primary CVE score

Vulnerability scoring details

Base CVSS 3.1 score

5.4Medium
CVSS 3.1 vector shape for CVE-2020-25154Attack VectorAttack ComplexityPrivileges RequiredUser InteractionScopeConfidentiality ImpactIntegrity ImpactAvailability Impact

Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N

Attack Vector
NetworkAdjacentLocalPhysical
Attack Complexity
LowHigh
Privileges Required
NoneLowHigh
User Interaction
NoneRequired
Scope
ChangedUnchanged
Confidentiality Impact
HighLowNone
Integrity Impact
HighLowNone
Availability Impact
HighLowNone
Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
B. Braun Melsungen AGSpaceComunspecified, unspecifiedListed
B. Braun Melsungen AGBattery pack with Wi-Fiunspecified, unspecifiedListed
B. Braun Melsungen AGData module compactplusA10, A11Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.