CVE-2019-10966: In GE Aestiva and Aespire versions 7100 and 7900, a vulnerability exists where serial devices are connected...

In GE Aestiva and Aespire versions 7100 and 7900, a vulnerability exists where serial devices are connected via an added unsecured terminal server to a TCP/IP network configuration, which could allow an attacker to remotely modify device configuration and silence alarms.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysis

Security readout for executives and security teams

This CVE affects GE Aestiva and Aespire 7100 and 7900 anesthesia systems when serial devices are connected through an added unsecured terminal server to TCP/IP. In that configuration, a remote attacker could change device configuration and silence alarms, creating potential patient-safety and operational risk. Exposure is likely limited to environments using affected GE Aestiva or Aespire 7100/7900 systems with added terminal servers bridging serial connections to TCP/IP networks. Treat as high priority in clinical environments if these anesthesia systems are network-bridged. The concern is not broad internet software exposure, but possible remote interference with configuration and alarms on patient-care equipment. Mitigation focus: Identify affected GE Aestiva and Aespire 7100/7900 deployments.; Remove or disable unsecured terminal-server network exposure where possible.; Restrict terminal-server access to trusted clinical engineering networks only..

Prepared

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · medium confidence lookup

CWE-287: Credential and account abuse lookup

Authentication and credential weaknesses can make valid-account abuse and credential telemetry useful review starting points. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2019-10966 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
2Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Source materials

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
GEAestiva and Aespireversions 7100, versions 7900Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-287 · source CWE mapping

Improper Authentication