CVE-2018-7510: In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prio...

In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 4107600010.23, passwords are presented in plaintext in a file that is accessible without authentication.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysishigh

Security readout for executives and security teams

Plain-English summary

This flaw can expose passwords in readable form through an unauthenticated file in the BeaconMedaes TotalAlert Scroll Medical Air Systems web application. For healthcare facilities using affected versions, it creates a credible path to unauthorized access. The sources do not provide CVSS scoring or evidence of active exploitation.

Executive priority

Treat this as a high-priority healthcare operational technology exposure if affected systems are present. The immediate business risk is unauthorized access through exposed credentials, with potential operational impact depending on network placement and system privileges.

Technical view

CVE-2018-7510 is a CWE-256 plaintext password storage/exposure issue affecting BeaconMedaes TotalAlert Scroll Medical Air Systems web application versions before 4107600010.23. Passwords are presented in plaintext in a file reachable without authentication. No exploit details, CVSS vector, or KEV listing are provided in the source bundle.

Likely exposure

Exposure is likely limited to organizations operating BeaconMedaes TotalAlert Scroll Medical Air Systems with the web application below version 4107600010.23, especially if reachable from broader hospital, vendor, or remote-access networks.

Exploitation context

The provided sources support unauthenticated access to plaintext passwords, but do not state that the vulnerability is being actively exploited. CISA KEV status is false in the bundle, so urgency is driven by credential exposure in a medical facility context, not confirmed exploitation.

Researcher notes

The bundle lacks CVSS detail, exploit status, and full vendor remediation text. Analysis should stay constrained to unauthenticated plaintext password exposure in affected versions. Avoid assuming device control impact without environment-specific access, privilege, and network segmentation evidence.

Mitigation direction

  • Upgrade affected systems to version 4107600010.23 or later if vendor guidance confirms availability.
  • Restrict web application access to trusted engineering or clinical operations networks only.
  • Remove internet or broad internal network reachability for affected interfaces.
  • Rotate any passwords that may have been exposed after remediation is complete.
  • Review BeaconMedaes and ICS-CERT guidance before operational changes.

Validation and detection

  • Inventory TotalAlert Scroll Medical Air Systems and record web application software versions.
  • Confirm whether any system runs a version earlier than 4107600010.23.
  • Check whether the web application is reachable from untrusted or remote-access networks.
  • Review access logs for unusual web requests or authentication activity.
  • Validate credential rotation after updating or otherwise remediating affected systems.
Prepared
Confidence
medium
Sources
3

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-256: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2018-7510 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
2Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Source materials

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
ICS-CERTBeaconMedaes TotalAlert Scroll Medical Air Systems web applicationAll versions prior to version 4107600010.23Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-256 · source CWE mapping

Plaintext Storage of a Password

Plaintext Storage of a Password represents a recurring weakness pattern that can create exploitable paths when design, validation, or implementation controls are missing.