LiveActive security incident?Get immediate response
CVE Record

CVE-2017-9658: Certain 802.11 network management messages have been determined to invoke wireless access point blacklistin...

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysis

Security readout for executives and security teams

This affects Philips IntelliVue MX40 patient-worn monitors using WLAN before B.06.18. Certain 802.11 management messages can make the monitor blacklist an access point unnecessarily, disrupting central monitoring until it reconnects elsewhere or staff reset it. The concern is patient-monitoring availability, not data theft. Exposure is limited to Philips IntelliVue MX40 Patient Worn Monitor WLAN-only deployments running versions before B.06.18. Risk is most relevant in hospital wireless environments where central monitoring depends on stable Wi-Fi association. Treat this as a patient-monitoring availability issue requiring planned remediation, especially in clinical areas relying on wireless MX40 monitoring. It is not described as data compromise, but interruption of central monitoring can create operational risk and staff intervention requirements. Mitigation focus: Inventory Philips IntelliVue MX40 WLAN-only monitors and confirm installed software versions.; Upgrade affected devices to Philips IntelliVue MX40 Version B.06.18 or later guidance-supported software.; Confirm disconnect messaging and alarming are enabled after the Philips update..

Prepared

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-755: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2017-9658 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
3Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
PhilipsIntelliVue MX40 Patient Worn MonitorIntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-755 · source CWE mapping

Improper Handling of Exceptional Conditions

Improper Handling of Exceptional Conditions represents a recurring weakness pattern that can create exploitable paths when design, validation, or implementation controls are missing.