CVE-2017-12714: Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of co...

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysis

Security readout for executives and security teams

Some Abbott pacemakers made before August 28, 2017 could be repeatedly awakened over nearby RF communication, consuming battery power faster than intended. This is a patient-safety and availability issue, not a data theft issue. Abbott developed a firmware update to help mitigate the vulnerability. Exposure is limited to listed Abbott pacemaker models manufactured before August 28, 2017. Organizations without those implanted or managed devices are not affected based on the provided sources. Treat as moderate urgency with clinical coordination. The issue affects device battery availability and patient safety, but the cited attack requires nearby access and high complexity, and no active exploitation is documented in the provided sources. Mitigation focus: Identify affected Abbott pacemaker models and manufacturing dates.; Follow Abbott firmware update guidance for affected devices.; Coordinate remediation through cardiology or electrophysiology care teams..

Prepared

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-920: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2017-12714 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
2Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Source materials

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
Abbott LaboratoriesAccent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.All versions of pacemakers manufactured prior to August 28, 2017Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.