LiveActive security incident?Get immediate response
CVE Record

CVE-2015-3952: Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3...

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysismoderate

Security readout for executives and security teams

Plain-English summary

Some Hospira infusion pumps stored wireless network keys in plain text. In a healthcare environment, this can turn a device-management weakness into a network access concern. The source bundle does not provide a CVSS score or evidence of active exploitation.

Executive priority

Treat this as a healthcare device hygiene issue with network security implications. Prioritize inventory and port closure for affected pumps, then confirm vendor-supported remediation or replacement options.

Technical view

CVE-2015-3952 is a CWE-312 cleartext storage issue affecting Hospira Plum A+ <=13.4, Plum A+3 <=13.6, and Symbiq <=3.13. Hospira recommended closing FTP Port 20 and TELNET Port 23 on affected devices. Plum 360 is stated as not vulnerable.

Likely exposure

Exposure is most likely in healthcare environments still operating the named Hospira infusion systems at or below the affected versions, especially where device FTP or Telnet access remains reachable on clinical or wireless networks.

Exploitation context

The provided sources do not show CISA KEV listing or active exploitation. They indicate stored wireless keys are exposed in plain text, but do not describe required access level, exploit maturity, or observed attacks.

Researcher notes

The bundle supports cleartext wireless key storage and specific affected versions, but lacks CVSS details, patch identifiers, exploit prerequisites, and exploitation evidence. Avoid assuming remote exploitation beyond the cited FTP/Telnet mitigation guidance.

Mitigation direction

  • Close Port 20/FTP on affected devices as Hospira recommended.
  • Close Port 23/TELNET on affected devices as Hospira recommended.
  • Identify whether any affected models remain in service.
  • Check Hospira or successor vendor guidance for supported remediation options.
  • Consider Plum 360 status when evaluating replacement options.

Validation and detection

  • Inventory Plum A+, Plum A+3, and Symbiq devices by model and version.
  • Confirm Plum A+ devices are above version 13.4 or remediated.
  • Confirm Plum A+3 devices are above version 13.6 or remediated.
  • Confirm Symbiq devices are above version 3.13 or remediated.
  • Verify network controls block FTP and Telnet access to affected devices.
Prepared
Confidence
medium
Sources
3

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-312: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2015-3952 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
2Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Source materials

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
HospiraPlum A+ Infusion System<= 13.4Listed
HospiraPlum A+3 Infusion System<= 13.6Listed
HospiraSymbiq Infusion System<= 3.13Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-312 · source CWE mapping

Cleartext Storage of Sensitive Information

Cleartext Storage of Sensitive Information represents a recurring weakness pattern that can create exploitable paths when design, validation, or implementation controls are missing.