LiveActive security incident?Get immediate response
CVE Record

CVE-2014-5431: Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) vers...

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

UnknownCVSS not scoredNot KEV-listedUpdated
Glexia's TakeAutomated analysismoderate

Security readout for executives and security teams

Plain-English summary

This issue affects a specific Baxter SIGMA Spectrum infusion system configuration. A built-in password could let someone with physical access change limited management settings, including wireless behavior and an audible phase-complete alarm. The bundle does not show remote exploitation, active exploitation, or a CVSS score.

Executive priority

Treat as a targeted medical-device exposure, not an internet-scale emergency. Prioritize inventory, physical access control, and vendor-supported upgrade planning for affected pumps.

Technical view

CVE-2014-5431 is a CWE-259 hard-coded password issue in Baxter SIGMA Spectrum Infusion System 6.05, model 35700BAX, with WBM version 16. Access is described as requiring physical device access and exposing basic biomedical information, limited settings, and WBM network configuration.

Likely exposure

Exposure appears limited to healthcare environments still operating Baxter SIGMA Spectrum version 6.05, model 35700BAX, with WBM version 16 connected.

Exploitation context

The source bundle says physical access is required. KEV status is false, and the provided sources do not document active exploitation or public weaponization.

Researcher notes

Evidence is narrow: one affected configuration, hard-coded password, physical access requirement, and limited management impact. No CVSS vector, exploit evidence, or detailed remediation steps are included in the bundle beyond Baxter version 8 availability.

Mitigation direction

  • Inventory SIGMA Spectrum devices and identify version 6.05 model 35700BAX with WBM version 16.
  • Assess migration to Baxter SIGMA Spectrum version 8 with Baxter support.
  • Restrict unsupervised physical access to affected infusion systems.
  • Review Baxter and ICS-CERT guidance before changing clinical configurations.

Validation and detection

  • Confirm device software version, model number, and WBM version from asset records.
  • Verify whether the wireless battery module is connected on each device.
  • Review WBM network and limited management settings for unauthorized changes.
  • Check clinical areas for physical access controls around affected devices.
Prepared
Confidence
medium
Sources
3

Generated from the cited source records. This long-tail analysis has not been individually reviewed by a named human.

Potential ATT&CK relevance

Conservative CVE-to-ATT&CK context

These mappings and lookup hints may be relevant to the vulnerability behavior, CWE, affected product, or exposure path. Glexia-inferred context is not an official MITRE, ATT&CK, CWE, or CVE Program mapping.

ATT&CK lookup starting points

Use these exact CWE pages and searches to review the Glexia ATT&CK library from this CVE's weakness and description context.

cwe · low confidence lookup

CWE-259: Exact CWE lookup

Use the exact CWE identifier as the starting point before reviewing related ATT&CK behavior. Open the exact CWE lookup page first, then review the ATT&CK searches from that MITRE weakness context. This is a Glexia lookup hint, not an official ATT&CK mapping.

Open ATT&CK lookup
description · low confidence lookup

Credential and access behavior lookup

The CVE wording references authentication or credential exposure, so valid-account and credential-access review may help. This is a Glexia inferred lookup path, not an official MITRE, ATT&CK, or CVE Program mapping.

Open ATT&CK lookup
cve · low confidence lookup

CVE-2014-5431 mapping review

Open the CVE-to-ATT&CK bridge for reviewed, inferred, or future official mappings tied to this CVE.

Open ATT&CK lookup
Vulnerability profileCVE Program record
Severity
Unknown
CVSS
Not scored
Known Exploited
No
Published
Official CVE source material

CNA and ADP enrichment extracted from CVE v5

These fields come from the CVE record and ADP containers, not from Glexia's Take. They preserve time-varying source decisions such as CISA SSVC, KEV status, CVSS metrics, and provider references.

0CVSS vectors
0Timeline events
0ADP providers
2Source links

CVSS and timeline data

No CVSS vectors or timeline events were available in the normalized CVE source material.

Source materials

Affected products

Products and packages named in the record

VendorProductVersion / packageStatus
BaxterSIGMA Spectrum Infusion System6.05 (model 35700BAX) with wireless battery module (WBM) version 16Listed
Weakness

CWE details

CWE links open Glexia weakness intelligence pages with official CWE context, developer remediation guidance, and related CVE mappings.

CWE-259 · source CWE mapping

Use of Hard-coded Password

Use of Hard-coded Password represents a recurring weakness pattern that can create exploitable paths when design, validation, or implementation controls are missing.