{"dataType":"CVE_RECORD","dataVersion":"5.2","cveMetadata":{"cveId":"CVE-2019-25722","assignerOrgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","state":"PUBLISHED","assignerShortName":"VulnCheck","dateReserved":"2026-06-02T18:39:44.176Z","datePublished":"2026-06-02T18:44:11.815Z","dateUpdated":"2026-06-03T16:19:46.284Z"},"containers":{"cna":{"providerMetadata":{"orgId":"83251b91-4cc7-4094-a5c7-464a1b83ea10","shortName":"VulnCheck","dateUpdated":"2026-06-03T16:19:46.284Z"},"title":"Dräger SC Monitoring Devices Hard-coded Credentials and DoS","datePublic":"2019-11-21T00:00:00.000Z","problemTypes":[{"descriptions":[{"lang":"en","cweId":"CWE-798","description":"CWE-798 Use of Hard-coded Credentials","type":"CWE"}]}],"affected":[{"vendor":"Dräger","product":"SC 6002XL","versions":[{"status":"affected","version":"SC 6002XL","versionType":"custom"}],"defaultStatus":"unknown"},{"vendor":"Dräger","product":"SC6802XL","versions":[{"status":"affected","version":"SC6802XL","versionType":"custom"}],"defaultStatus":"unknown"},{"vendor":"Dräger","product":"SC 7000","versions":[{"status":"affected","version":"SC 7000","versionType":"custom"}],"defaultStatus":"unknown"},{"vendor":"Dräger","product":"SC8000","versions":[{"status":"affected","version":"SC8000","versionType":"custom"}],"defaultStatus":"unknown"},{"vendor":"Dräger","product":"SC90000 XL","versions":[{"status":"affected","version":"SC90000 XL","versionType":"custom"}],"defaultStatus":"unknown"}],"descriptions":[{"lang":"en","value":"Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.","supportingMedia":[{"type":"text/html","base64":false,"value":"<p>Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.</p>"}]}],"references":[{"url":"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf","tags":["vendor-advisory"]},{"url":"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-hard-coded-credentials-and-dos","tags":["third-party-advisory"]}],"metrics":[{"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}],"cvssV4_0":{"attackVector":"ADJACENT","attackComplexity":"LOW","attackRequirements":"NONE","privilegesRequired":"NONE","userInteraction":"NONE","vulnConfidentialityImpact":"LOW","subConfidentialityImpact":"NONE","vulnIntegrityImpact":"LOW","subIntegrityImpact":"NONE","vulnAvailabilityImpact":"HIGH","subAvailabilityImpact":"NONE","exploitMaturity":"NOT_DEFINED","Safety":"NOT_DEFINED","Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","valueDensity":"NOT_DEFINED","vulnerabilityResponseEffort":"NOT_DEFINED","providerUrgency":"NOT_DEFINED","version":"4.0","baseSeverity":"HIGH","baseScore":7.2,"vectorString":"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N"}},{"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}],"cvssV3_1":{"version":"3.1","attackVector":"ADJACENT_NETWORK","attackComplexity":"LOW","privilegesRequired":"NONE","userInteraction":"NONE","scope":"UNCHANGED","confidentialityImpact":"LOW","integrityImpact":"LOW","availabilityImpact":"HIGH","baseSeverity":"HIGH","baseScore":7.6,"vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H"}}],"credits":[{"lang":"en","value":"Jeroen Slobbe and Max Grim","type":"finder"}],"source":{"discovery":"EXTERNAL"},"x_generator":{"engine":"Vulnogram 1.0.2"}},"adp":[{"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2026-06-03T13:15:43.146803Z","id":"CVE-2019-25722","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"partial"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2026-06-03T13:15:56.970Z"}}]}}